Precision Tools for Visual Disorder Detection
Visiolite® GT
Empowering Vision Testing with Innovation
|
The Visiolite® GT is an advanced vision screening device designed for occupational health services and preventive medicine specialists. Featuring customizable test sequences, a patented motorized system for comprehensive visual field analysis, and intuitive software, it ensures precise assessments of visual acuity and disorders. Its ergonomic design, sensory sensors, and options for computerized testing make the Visiolite® GT an essential tool for modern vision care, combining simplicity with efficiency.
Visiolite® GT
- Features
- Technical Specs
- Brochures
- Manual
- CE certifications
- Videos
- Far vision lenses
- Near vision lenses
- Removable forehead rest piece – A removable elastomer piece is positioned on the forehead rest. An electro-sensitive sensor indicates the patient’s forehead is in contact with the forehead rest, via a Visiolite® light display. When the forehead is correctly positioned on the forehead rest, the patient should not feel any discomfort.
- Peripheral field – A number of small holes are visible around the lenses. These light guides are for the visual peripheral field test.
- Nose position – A light on the bottom right of the screen indicates whether the Visiolite® is correctly connected:
The Visiolite® Essential remote control is ergonomic and very user-friendly. Each key of the remote control corresponds to a test, and each key is associated with a light for the test displayed in the Visiolite®.
The right side of the remote control is for visual acuity tests. The left part is for extra tests.
To scroll through the tests of the selected mode, use the left and right arrows. Go from one mode to another by pressing on the right or left arrow.
Technical features | |
Storage temperature Operating temperature Humidity Altitude of operation External power plug Power supply Power absorbed Standard Medical class Software security class Electric class Applied part Marking Dimensions Weight | 0 – 50°C (32 – 122°F) |
15 – 35°C (59 – 95°F) | |
75% maximum | |
< 2000m (6561.7ft) | |
Input: 100-240VAC 50-60Hz 0.6A Output: 12VDC 2.08A (medical class) | |
12VDC from an external power medical supply (refer § 1.1) | |
24W | |
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, EN 62366-1, EN ISO 10993-5, EN ISO 10993- 10, NF EN ISO 14971, NF EN 62304/A1, NF EN 1041+A1, EN ISO 15223-1, ISO 8596, ANSI Z80.21, NF EN ISO 15004-2 :2007 §5.4.1.6, EN ISO 10993-1. | |
Class I | |
A | |
Class II | |
Type B | |
CE | |
50x27x25 cm (19.7×10.6×9.8 inches) | |
4.850 kg (10.7lb) |
Optical features | |
Lighting system Lenses Lighting conditions (nominal values) | 16 white LEDs and diffuser system |
Depending of the version: Far vision: (5.0 ± 0.1)m/(16.4 ± 0.3)ft ; (20.0 ± 0.4)ft Intermediate vision: (60.0 ± 0.5)cm/(23.6 ± 0.2)″ ; (80.0 ± 0.5)cm/(31.5± 0.2)″ ; (24.0 ± 0.2)″ Near vision: (33.00 ± 0.25)cm/(13.0 ± 0.1)″ ; (35.50 ± 0.25)cm/(14.0 ± 0.1)″ ; (16.0 ± 0.1)″ Hyperopia lenses: +1 dioptre | |
High photopic (160 candelas) Low photopic (80 candelas) Dusk mesopic (3 candelas) Conform to NF EN ISO 8596 standard |
Test features | |
Reactivity Average test time | Time between 2 neighboring tests: 700ms Time to pass from one slide to another: 1 sec |
Routine test: 3 mins In-depth test: 5 mins |
Essential Modulus Master | |||
Acuity Tests | |||
Landolt Rings | • | • | • |
Numbers | • | • | |
Letters | • | • | • |
Low vision letters | • | • | |
Supplementary tests | |||
Astigmatism | • | • | • |
Red/green duochrome | • | • | • |
Depth perception | • | • | |
Vertical & horizontal phorias | • | • | • |
Fusion | • | ||
Ishihara type colours test | • | • | • |
Amsler grid | • | • | |
Contrast sensivity | • | ||
Hyperopia test (+1 dioptre) | • | • | • |
Horizontal & vertical visual field | • | • | • |
Central visual field | • | • | |
Glare sensitivity | • (GT option) | ||
Glare resistance (educational test) | • (GT option) | ||
Distances | |||
Far vision 5m/16.4ft | • | • | • |
Intermediate vision 60cm/23.6inches | • | • | |
Near vision 33cm/13.0inches | • | • | • |
Lighting | |||
High photopic | • | • | • |
Low photopic | • | • | |
Mesopic (night vision) | • | • | |
Driver mode | |||
Computer | • | • | |
Remote control | • | • |